Quality Assurance of MRI Systems. Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2 Quality Control of MRI Systems



Σχετικά έγγραφα
the total number of electrons passing through the lamp.

Main source: "Discrete-time systems and computer control" by Α. ΣΚΟΔΡΑΣ ΨΗΦΙΑΚΟΣ ΕΛΕΓΧΟΣ ΔΙΑΛΕΞΗ 4 ΔΙΑΦΑΝΕΙΑ 1

ΕΙΣΑΓΩΓΗ ΣΤΗ ΣΤΑΤΙΣΤΙΚΗ ΑΝΑΛΥΣΗ

ΚΥΠΡΙΑΚΗ ΕΤΑΙΡΕΙΑ ΠΛΗΡΟΦΟΡΙΚΗΣ CYPRUS COMPUTER SOCIETY ΠΑΓΚΥΠΡΙΟΣ ΜΑΘΗΤΙΚΟΣ ΔΙΑΓΩΝΙΣΜΟΣ ΠΛΗΡΟΦΟΡΙΚΗΣ 19/5/2007

Instruction Execution Times

HOMEWORK 4 = G. In order to plot the stress versus the stretch we define a normalized stretch:

Math 6 SL Probability Distributions Practice Test Mark Scheme

Section 8.3 Trigonometric Equations

5.4 The Poisson Distribution.

Potential Dividers. 46 minutes. 46 marks. Page 1 of 11

6.003: Signals and Systems. Modulation

EE512: Error Control Coding

(1) Describe the process by which mercury atoms become excited in a fluorescent tube (3)

Surface Mount Multilayer Chip Capacitors for Commodity Solutions

Strain gauge and rosettes

Other Test Constructions: Likelihood Ratio & Bayes Tests

NMBTC.COM /

Surface Mount Aluminum Electrolytic Capacitors

[1] P Q. Fig. 3.1

Απόκριση σε Μοναδιαία Ωστική Δύναμη (Unit Impulse) Απόκριση σε Δυνάμεις Αυθαίρετα Μεταβαλλόμενες με το Χρόνο. Απόστολος Σ.

Approximation of distance between locations on earth given by latitude and longitude

Numerical Analysis FMN011

Πτυχιακή Εργασι α «Εκτι μήσή τής ποιο τήτας εικο νων με τήν χρή σή τεχνήτων νευρωνικων δικτυ ων»

EPL 603 TOPICS IN SOFTWARE ENGINEERING. Lab 5: Component Adaptation Environment (COPE)

Statistical Inference I Locally most powerful tests

Thin Film Chip Resistors

Calculating the propagation delay of coaxial cable

Homework 8 Model Solution Section

SCOPE OF ACCREDITATION TO ISO 17025:2005

ST5224: Advanced Statistical Theory II

ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΕΙΡΑΙΑ ΤΜΗΜΑ ΝΑΥΤΙΛΙΑΚΩΝ ΣΠΟΥΔΩΝ ΠΡΟΓΡΑΜΜΑ ΜΕΤΑΠΤΥΧΙΑΚΩΝ ΣΠΟΥΔΩΝ ΣΤΗΝ ΝΑΥΤΙΛΙΑ

Capacitors - Capacitance, Charge and Potential Difference

Phys460.nb Solution for the t-dependent Schrodinger s equation How did we find the solution? (not required)

Second Order RLC Filters

Econ 2110: Fall 2008 Suggested Solutions to Problem Set 8 questions or comments to Dan Fetter 1

Copernicus for Local and Regional Authorities

CHAPTER 25 SOLVING EQUATIONS BY ITERATIVE METHODS

The Simply Typed Lambda Calculus

Démographie spatiale/spatial Demography

Μειέηε, θαηαζθεπή θαη πξνζνκνίσζε ηεο ιεηηνπξγίαο κηθξήο αλεκνγελλήηξηαο αμνληθήο ξνήο ΓΗΠΛΩΜΑΣΗΚΖ ΔΡΓΑΗΑ


ΚΥΠΡΙΑΚΗ ΕΤΑΙΡΕΙΑ ΠΛΗΡΟΦΟΡΙΚΗΣ CYPRUS COMPUTER SOCIETY ΠΑΓΚΥΠΡΙΟΣ ΜΑΘΗΤΙΚΟΣ ΔΙΑΓΩΝΙΣΜΟΣ ΠΛΗΡΟΦΟΡΙΚΗΣ 6/5/2006

Precision Metal Film Fixed Resistor Axial Leaded

1. Ηλεκτρικό μαύρο κουτί: Αισθητήρας μετατόπισης με βάση τη χωρητικότητα

ΚΥΠΡΙΑΚΗ ΕΤΑΙΡΕΙΑ ΠΛΗΡΟΦΟΡΙΚΗΣ CYPRUS COMPUTER SOCIETY ΠΑΓΚΥΠΡΙΟΣ ΜΑΘΗΤΙΚΟΣ ΔΙΑΓΩΝΙΣΜΟΣ ΠΛΗΡΟΦΟΡΙΚΗΣ 11/3/2006

b. Use the parametrization from (a) to compute the area of S a as S a ds. Be sure to substitute for ds!

SMD Transient Voltage Suppressors

Aluminum Electrolytic Capacitors (Large Can Type)

First Sensor Quad APD Data Sheet Part Description QA TO Order #

ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΤΜΗΜΑ ΗΛΕΚΤΡΟΛΟΓΩΝ ΜΗΧΑΝΙΚΩΝ ΚΑΙ ΤΕΧΝΟΛΟΓΙΑΣ ΥΠΟΛΟΓΙΣΤΩΝ ΤΟΜΕΑΣ ΣΥΣΤΗΜΑΤΩΝ ΗΛΕΚΤΡΙΚΗΣ ΕΝΕΡΓΕΙΑΣ

Statistics 104: Quantitative Methods for Economics Formula and Theorem Review

Aluminum Electrolytic Capacitors

ΤΕΧΝΟΛΟΓΙΚΟ ΠΑΝΕΠΙΣΤΗΜΙΟ ΚΥΠΡΟΥ ΤΜΗΜΑ ΝΟΣΗΛΕΥΤΙΚΗΣ

ΔΙΠΛΩΜΑΤΙΚΗ ΕΡΓΑΣΙΑ. του φοιτητή του Τμήματος Ηλεκτρολόγων Μηχανικών και. Τεχνολογίας Υπολογιστών της Πολυτεχνικής Σχολής του. Πανεπιστημίου Πατρών

department listing department name αχχουντσ ϕανε βαλικτ δδσϕηασδδη σδηφγ ασκϕηλκ τεχηνιχαλ αλαν ϕουν διξ τεχηνιχαλ ϕοην µαριανι

Monolithic Crystal Filters (M.C.F.)

PARTIAL NOTES for 6.1 Trigonometric Identities

Multilayer Ceramic Chip Capacitors

Correction Table for an Alcoholometer Calibrated at 20 o C

EE101: Resonance in RLC circuits

Code Breaker. TEACHER s NOTES

Jesse Maassen and Mark Lundstrom Purdue University November 25, 2013

2R2. 2 (L W H) [mm] Wire Wound SMD Power Inductor. Nominal Inductance Packing Tape & Reel. Design Code M ±20%

Reminders: linear functions

ΓΕΩΜΕΣΡΙΚΗ ΣΕΚΜΗΡΙΩΗ ΣΟΤ ΙΕΡΟΤ ΝΑΟΤ ΣΟΤ ΣΙΜΙΟΤ ΣΑΤΡΟΤ ΣΟ ΠΕΛΕΝΔΡΙ ΣΗ ΚΤΠΡΟΤ ΜΕ ΕΦΑΡΜΟΓΗ ΑΤΣΟΜΑΣΟΠΟΙΗΜΕΝΟΤ ΤΣΗΜΑΣΟ ΨΗΦΙΑΚΗ ΦΩΣΟΓΡΑΜΜΕΣΡΙΑ

Repeated measures Επαναληπτικές μετρήσεις

Multilayer Ceramic Chip Capacitors

ΠΑΝΔΠΗΣΖΜΗΟ ΠΑΣΡΩΝ ΣΜΖΜΑ ΖΛΔΚΣΡΟΛΟΓΩΝ ΜΖΥΑΝΗΚΩΝ ΚΑΗ ΣΔΥΝΟΛΟΓΗΑ ΤΠΟΛΟΓΗΣΩΝ ΣΟΜΔΑ ΤΣΖΜΑΣΩΝ ΖΛΔΚΣΡΗΚΖ ΔΝΔΡΓΔΗΑ

High Performance Voltage Controlled Amplifiers Typical and Guaranteed Specifications 50 Ω System

UDZ Swirl diffuser. Product facts. Quick-selection. Swirl diffuser UDZ. Product code example:

derivation of the Laplacian from rectangular to spherical coordinates

Εργαστήριο Ανάπτυξης Εφαρμογών Βάσεων Δεδομένων. Εξάμηνο 7 ο

Modbus basic setup notes for IO-Link AL1xxx Master Block

4.6 Autoregressive Moving Average Model ARMA(1,1)

ΑΝΙΧΝΕΥΣΗ ΓΕΓΟΝΟΤΩΝ ΒΗΜΑΤΙΣΜΟΥ ΜΕ ΧΡΗΣΗ ΕΠΙΤΑΧΥΝΣΙΟΜΕΤΡΩΝ ΔΙΠΛΩΜΑΤΙΚΗ ΕΡΓΑΣΙΑ

Solutions to Exercise Sheet 5

Block Ciphers Modes. Ramki Thurimella

Προσομοίωση BP με το Bizagi Modeler

LS series ALUMINUM ELECTROLYTIC CAPACITORS CAT.8100D. Specifications. Drawing. Type numbering system ( Example : 200V 390µF)

Smaller. 6.3 to 100 After 1 minute's application of rated voltage at 20 C, leakage current is. not more than 0.03CV or 4 (µa), whichever is greater.

Estimation for ARMA Processes with Stable Noise. Matt Calder & Richard A. Davis Colorado State University

Matrices and Determinants

Daewoo Technopark A-403, Dodang-dong, Wonmi-gu, Bucheon-city, Gyeonggido, Korea LM-80 Test Report

Μηχανική Μάθηση Hypothesis Testing

ΠΑΡΑΜΕΤΡΟΙ ΕΠΗΡΕΑΣΜΟΥ ΤΗΣ ΑΝΑΓΝΩΣΗΣ- ΑΠΟΚΩΔΙΚΟΠΟΙΗΣΗΣ ΤΗΣ BRAILLE ΑΠΟ ΑΤΟΜΑ ΜΕ ΤΥΦΛΩΣΗ

2 Composition. Invertible Mappings

C.S. 430 Assignment 6, Sample Solutions

SCHOOL OF MATHEMATICAL SCIENCES G11LMA Linear Mathematics Examination Solutions

«Χρήσεις γης, αξίες γης και κυκλοφοριακές ρυθμίσεις στο Δήμο Χαλκιδέων. Η μεταξύ τους σχέση και εξέλιξη.»

ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΤΜΗΜΑ ΗΛΕΚΤΡΟΛΟΓΩΝ ΜΗΧΑΝΙΚΩΝ ΚΑΙ ΤΕΧΝΟΛΟΓΙΑΣ ΥΠΟΛΟΓΙΣΤΩΝ ΤΟΜΕΑΣ ΣΥΣΤΗΜΑΤΩΝ ΗΛΕΚΤΡΙΚΗΣ ΕΝΕΡΓΕΙΑΣ

ΠΟΛΥΤΕΧΝΕΙΟ ΚΡΗΤΗΣ ΣΧΟΛΗ ΜΗΧΑΝΙΚΩΝ ΠΕΡΙΒΑΛΛΟΝΤΟΣ

Reaction of a Platinum Electrode for the Measurement of Redox Potential of Paddy Soil

ΜΕΛΕΤΗ ΚΑΙ ΠΡΟΣΟΜΟΙΩΣΗ ΙΑΜΟΡΦΩΣΕΩΝ ΣΕ ΨΗΦΙΑΚΑ ΣΥΣΤΗΜΑΤΑ ΕΠΙΚΟΙΝΩΝΙΩΝ.

TMA4115 Matematikk 3

MSM Men who have Sex with Men HIV -

Concrete Mathematics Exercises from 30 September 2016

Metal Oxide Varistors (MOV) Data Sheet

Test Data Management in Practice

ΤΕΙ ΚΑΒΑΛΑΣ ΣΧΟΛΗ ΤΕΧΝΟΛΟΓΙΚΩΝ ΕΦΑΡΜΟΓΩΝ ΤΜΗΜΑ ΗΛΕΚΤΡΟΛΟΓΙΑΣ ΠΤΥΧΙΑΚΗ ΕΡΓΑΣΙΑ

ECE 468: Digital Image Processing. Lecture 8

Transcript:

Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Quality Control of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Objectives To gain a general understanding of MR system variability To be able to set up and manage an MRI QC and Compliance Testing program To gain a general understanding of MRI Artifacts 1

Slide 4 Objectives of an MRI QC Program To implement a system practical enough for technologists to utilize on a daily basis To obtain meaningful results which reveal system changes before patient care is affected To document problems and corrective action in a manner which satisfies accreditation and regulatory requirements Slide 5 System Variability Magnetic resonance imaging systems are subject to variability from: Drift in RF Electronics RF Coils RF Transceiver Chain Magnetic Field Decay Foreign ferro- or para- magnetic material Introduction of material producing MRI signal Gradient system failures Slide 6 Effects of System Variability Radio Frequency effects SNR Loss Occurs when the RF center frequency is off resonance Power is insufficient to produce 90 or 180 deg pulses If the RF coil is detuned If the RF coil homogeneity is not optimized Noise Can increase with failures in preamplifier or amplifier Can be introduced with failures in the RF shielding (most often door). 2

Slide 7 Gradient effects Geometric Distortion Can occur with when maximum amplitude is not achieved or gradient is miscalibrated. Can occur if eddy current compensation changes Spatially Dependent SNR Loss Can occur when local field homogeneity decreases Can occur when gradient waveforms are not optimized (eg eddy currents) resulting in incomplete rephasing of echo Phase Errors Cause misregistration in Phase Encoding direction Can arise from any of the 3 orthogonal gradients Slide 8 Magnetic Field Inhomogeneity SNR Loss Can occur when local total field homogeneity decreases T2*. Can occur when center frequency is shifted beyond bandwidth of receiver In Plane Geometric Distortion Can cause regions of low or high signal when resonant frequency is shifted beyond bandwidth of a pixel Slide 9 Magnetic Field Inhomogeneity Geometric Distortion in Slice Select Direction Can cause misregistration of signal as a function of slice location, or distortion of slice profile causing increased crosstalk. Chemical Shift Frequency Offsets Frequency selective pulses (e.g. lipid or water saturation, mtc) can have a spatial dependence to their intended effect. 3

Slide 10 Daily Testing ACR MRI Accreditation Recommendation Magnetic Field Stability (Center Frequency) Signal to Noise Ratio Artifact Inspection Slide 11 Standard Pulse Sequences T1 Weighted TR 500 2000 T2 Weighted TE min full <20msec 20/80 or 30/90 Matrix 128 or 160 128 or 160 Coil Head Head #excitations 1 1 FOV 24or 25 24 or 25 Slide 12 Magnetic Field Stability (Center Frequency) Test Object Procedure Sequence Parameters What to record Factors influencing measurement 4

Slide 13 Center Frequency (cont( Typical magnetic field stability is on the order of 0.1 ppm per hour or 2.4 ppm per day. At 1T this corresponds to approximately 100 Hz decay of the center frequency per day and 150 Hz per day at 1.5 tesla. A graph of center frequency vs. time should indicate steady decay, however, typical inhomogeneity over a 20 cm DSV is also within this range. Therefore, it is important to set up conditions over which reproducible measurements of the center frequency can be obtained. Slide 14 Center Frequency (cont( Additional magnetic fields due to variation in magnetic susceptibility can distort the line shape of the resultant water signal and produce an error in determining the peak position. This is minimized by using a spherically symmetric uniform phantom. Measurement of center frequency is further complicated by the presence of additional static magnetic fields used to adjust the homogeneity for specific applications (shims). The shim settings can be modified in two ways. Slide 15 Center Frequency (cont( Routine preventative maintenance for MRI calls for periodic re-shimming of the magnet using all the available set of shim fields. These shims involve coils designed to vary magnetic field as a function of spatial coordinates. Usually the set of coils includes shims which perturb the magnetic field spanning geometry characterized by second order spherical harmonics. These shim settings are usually not changed by the user. 5

Slide 16 Center Frequency (cont( Many manufacturers provide user controlled shimming using the three linear gradient magnetic fields, either manually or with a computer controlled algorithm (autoshimming). Some manufacturers utilize autoshimming before each pulse sequence. Slide 17 Center Frequency (cont( cont) Unaccounted shimming can add an additional magnetic field resulting in an increase in the center frequency from day to day. It is important therefore to use the same setting of the shim magnetic field gradients for each measurement. The graph of center frequency over a period of 15 months shows an increase in center frequency at several time points. The overall decrease in magnetic field is less than 160 Hz << less than manufacturer s specification Slide 18 Center Frequency over 15 Month Period 63860900 63860850 63860800 63860750 63860700 63860650 63860600 7/7/94 7/18/94 8/1/94 8/15/94 8/29/94 1/16/95 2/20/95 4/25/95 5/8/95 8/7/95 9/11/95 9/26/95 6

Slide 19 Center Frequency (cont( cont) Procedure Prescribe a standard T1 weighted single slice pulse sequence Make sure any auto shimming algorithm is turned off. Set up parameters for the auto pre-scan and load the preset values of the shims. Perform the clinical auto pre-scan procedure. Slide 20 Center Frequency (cont( cont) Procedure Record the time of the measurement, resultant center frequency, and shim file settings. Optionally record the transmitter and receiver gains. If the same shim file setting was used subtract the previous days frequency from the current value and compare with the specified decay. If the measurements were not acquired at the same time on successive days, adjust for total decay time. Action level: decay within manufacturer specification. Slide 21 Signal to Noise Ratio Test Objects Material Geometry RF Homogeneity Magnetic Susceptibility Effects Positioning Coil 7

Slide 22 Signal to Noise Ratio In measuring signal to noise ratio, a uniform test object is necessary in order to minimize loss of signal due to variation in magnetic susceptibility as well as spatial variation in the signal intensity due to RF inhomogeneity. Solutions which are not comparable to tissue conductivity may not load the coil properly. Slide 23 Signal to Noise Ratio (cont( The choice of coil is determined by the largest volume of procedures performed clinically, the best homogeneity characteristics, and the ease of both acquiring and analyzing the data. Though in some instances the head coil is not always used for the largest number of procedures, it usually has superior homogeneity compared to surface coils used in spine or extremity imaging. Slide 24 Signal to Noise Ratio (cont( Example Pulse sequence TR - 500 to 800 milliseconds TE - 10 to 20 milliseconds Single 10mm slice at isocenter. Single excitation, 256 x 160 matrix FOV - 20 to 24cm. 8

Slide 25 Signal to Noise Ratio (cont( These parameters are used to generate a T1 weighted image. Specific TR, TE etc., can be taken from a common clinical protocol, provided the T1 of the test object material is within the range of normal tissue. Once chosen, these parameters should be remained fixed as they all affect the resultant signal to noise ratio. Slide 26 Signal to Noise Ratio (cont( Use of the minimum TE as calculated by the imaging system is not recommended since this value will change with changes to software and hardware. Test object dimension and FOV should be chosen to fill greater than 85% of the usable field view of the coil in order to adequately take into account RF homogeneity effects. Slide 27 Signal to Noise Ratio (cont( ACR Accreditation T1 Series TR 500 TE Matrix 256 Coil #excitations 1 20 msec Head FOV 24or 25 Time 2:16 Use of 256 matrix is not necessary for SNR, however, if using the ACR phantom, other tests may be analyzed. The extra minute is small compared to set up time (aligning the phantom). 9

Slide 28 Signal to Noise Ratio (cont( Noise Characteristics Theoretically, noise in magnetic resonance is random. In practice however, there can be systematic contributions to background noise arising mainly from unwanted phase shifts acquired in the RF transceiver chain. Since phase is used in the image reconstruction algorithm to supply spatial information, these phase shifts produce spatial variation in this background signal ( ghosts ). Slide 29 Signal to Noise Ratio In measuring signal to noise ratio, a uniform test object is necessary in order to minimize loss of signal due to variation in magnetic susceptibility as well as spatial variation in the signal intensity due to RF inhomogeneity. Solutions which are not comparable to tissue conductivity may not load the coil properly. Slide 30 Analysis If S is the mean of the ROI from the original image and σ N is the standard deviation of an ROI from the noise image, then S SNR = 2 σ N Record the SNR along with means and standard deviations. Action level: SNR within manufacturer specification 10

Slide 31 Signal to Noise Ratio (cont( Noise Estimation: Method 1 While it may be attractive to measure the noise of a region of interest from a non signal producing area of the image field of view outside the test object using a single image, this practice may introduce systematic noise into the calculation of the SNR. If a single acquisition is desired, estimate noise from multiple regions in the image. Slide 32 Signal to Noise Ratio (cont( Procedure 1. Scan test object once using T1 series. 2. Choose ROI that encompasses at least 75% of the test object. Measure and record mean and standard deviation of the ROI 3. Calculate PSG from ACR accreditation. 4. If PSG acceptable, use average of 4 ROI σ s to estimate noise. Benefit is that PSG measurement can be tracked Slide 33 Signal to Noise Ratio (cont( Susceptibility from air bubble ACR T1 Series Slice #7 ROIs for systematic noise measurement Signal ROI 11

Slide 34 Signal to Noise Ratio (cont( SNR and PSG for three systems Fixed 11.5T Fixed 2 1.5T Mobile 1.0 T Mean σ Mean σ Mean σ 12.08 5.82 11.22 5.87 20.96 11.77 L R 11.18 6.87 10.59 6.15 22.59 10.40 T 10.42 5.13 9.93 5.07 20.28 10.18 B 9.04 4.41 10.47 5.14 19.14 10.04 Signal 1168.00 50.35 1407.19 40.32 1451.46 50.00 Noise Ave 10.68 5.56 10.55 5.56 20.74 10.60 PSG 0.001627 0.000501 0.001423 SNR 297.2202 358.0868 193.6942 Slide 35 Signal to Noise Ratio (cont( Noise Estimation: Method 2 An accepted practice is to image the test object twice, repeating the acquisition within a few minutes of the first. The noise is estimated from an ROI in an image formed by subtracting the first acquisition from the second. The assumption is that the signal (and noise) in the two images are uncorrelated. Slide 36 Noise Image 12

Slide 37 Signal to Noise Ratio (cont( Procedure 1. Scan test object twice using same parameters. 2. Subtract the second image from the first on a pixel by pixel basis. (This may require the images to be downloaded to a work station). 3. Choose ROI that encompasses at least 75% of the test object. Measure and record mean and standard deviation of the ROI in both the original images and the subtracted noise image. Slide 38 SNR Method 2 1.5T Body Coil Plane S σ SNR Axia l 1185 11.5 145.7255 Sa gitta l 1184 11.62 144.0989 Corona l 1174 11.5 144.3728 Slide 39 SNR over an 15 Month Period 110 108 106 104 102 100 98 96 94 92 90 7/7/94 7/18/94 8/1/94 8/15/94 8/29/94 1/16/95 2/20/95 4/25/95 5/8/95 8/7/95 9/11/95 9/26/95 A graph of SNR in the head coil over 15 months show day to day variation of SNR with a mean =104.3 and σ=2.6 13

Slide 40 Systematic Background Axial Body Coil FOV 36 cm Slide 41 Systematic Background Sagittal Coronal Slide 42 Weekly Testing ACR Processor sensitometry Other Phase Stability Magnetic Field Homogeneity Signal - to - Noise Ratio in orthogonal planes Image Uniformity 14

Slide 43 Processor Sensitometry Though analogous to other digital imaging modalities, printing of MR films can be problematic due to variation in the output of the MR system. This variation is due to Variable gain in the transceiver subsystem Variation in tissue contrast (T1, T2, ) Variation in operator selectable parameters Slide 44 Processor Sensitometry Example Protocol Display SMPTE pattern Inspect SMPTE image on the monitor Note 5% and 95% levels visibility Note any artifacts Film image using preset window and level Window width 256, Window level 128 Scan output image and record Speed, Contrast, Base+Fog Slide 45 Modified SMPTE Image 15

Slide 46 Tone Scale Calibration Optical De ns ity 3.00 2.50 2.00 1.50 1.00 0.50 0.00 1 4 7 10 13 16 19 Density Step Slide 47 Processor Sensitometry This procedure allows a standardized way of comparing the image on the monitor with the output of the laser printer. The film densities may be scanned, plotted, and interpreted in a manner consistent with other processors used in the department, however, this test does not narrow the source of a potential problem to either the monitor, MRI console output (either video or digital), laser printer performance, or processor performance. Slide 48 Processor QC 16

Slide 49 125 120 115 SNR Results mean =104.3 and σ=2.6 +/- 10% 110 105 100 95 90 85 80 75 7/7/94 7/18/94 8/1/94 8/15/94 8/29/94 1/16/95 2/20/95 4/25/95 5/8/95 8/7/95 9/11/95 9/26/95 Slide 50 Daily Change in Frequency 1.5T 1.0T υ 200 150 100 50 0-50 -100-150 -200 7/7/94 7/18/94 8/1/94 8/15/94 8/29/94 1/16/95 2/20/95 4/25/95 5/8/95 8/7/95 9/11/95 9/26/95 Slide 51 Review of Test Results Technologist - Daily Checklist Inspect QC phantom for artifacts Inspect SMPTE Pattern for artifacts (monitor) Inspect SMPTE Pattern for artifacts (film) Graph Center Frequency, SNR, Sensitometry Physicist Review of Daily QC Monthly or Quarterly (ACR Semi-Annual) Review Artifacts, Action Limits -ad hoc 17

Slide 1 Artifact Inspection MRI is susceptible to artifacts: Low SNR (and low energy) compared to other imaging modalities Ability to directly modify spatial frequency sampling Sensitivity to metallic objects Patient RF coil interaction is variable Slide 2 Artifact Inspection Artifacts may appear differently in test objects and patients because of differences in: Magnetic susceptibility Coil loading T1, T2 of test object material High contrast between plastic (no signal) and fluid Slide 3 EPI 1/2 FOV Phase Artifact 18

Slide 4 Artifact Inspection A single source may cause artifacts to appear differently Using different coils Using different pulse sequence parameters Anatomic plane pulse sequence type (eg spin echo vs. gradient echo) timing (TR, TE, gating) Spatial resolution (FOV, matrix size) Using different materials Slide 5 Gradient Echo vs. SE of Bleed Slide 6 RF Artifacts External RF Noise Specific frequency usually one pixel in width Phase Incoherent Independent of anatomic plane May change position with FOV or bandwidth 19

Slide 7 External RF Noise Slide 8 External RF Noise Slide 9 RF Artifacts Transmit Receive switch failure Inductive coupling of transmit coil and surface coil Spatial variation in RF power 20

Slide 10 Transmit - Surface Coil Artifact Phased Array Spine Slide 11 ADC Overflow (Clipping) Occurs when the signal is larger than the maximum ADC value, usually if gain is set too high. Slide 12 Non Phase Encoded Signal Stimulated Echoes FID (inhomogeneous B1, gradients) Signal from outside FOV Usually if phase and frequency directions swapped. Small FOV in larger body part 21

Slide 13 Stimulated Echo Interference Fringe spacing inversely proportional to difference in echo timing. Slide 14 Axial, Sagittal Lumbar Spine Slide 15 Homogeneity Artifacts Local inhomogeneity causes shifts in position in frequency encoding direction causes dephasing within a voxel causes frequency selective pulses to be spatially shifted Generally independent of pulse sequence parameters except worse on gradient echo vs spin echo saturation pulses 22

Slide 16 Homogeneity Artifacts Shim Misadjusted SE GRE Slide 17 Homogeneity/Susceptibility Artifact Homogeneity Susceptibility Slide 18 Metallic Artifacts - SE Spine 23

Slide 19 Metallic Artifacts - FSE Spine Spatial Sat Slide 20 Metallic Artifacts - SE Spine Wire Removed Slide 21 Metallic Artifacts SE Knee Coronal 24

Slide 22 Metallic Artifacts FSE Knee Coronal Fat Sat Slide 23 Metallic Artifacts FSE Knee Sagittal Fat Sat Slide 24 Gradient Artifacts Geometric Distortion Similar to Metallic Artifact but usually over FOV Direction Dependent In plane Frequency vs. Phase Slice profile distortion 25

Slide 25 Geometric Distortion Slide 26 Phase Artifacts Gradient effects Spatially dependent SNR loss Ghosts Differentiate from motion Non gradient effects Phase Modulation Ghosts DC offset Slide 27 Phase Modulation Artifact Weak Ghost displaced in PE direction Periodicity dependent upon synchronization with TR Line voltage, harmonics Gradient instability 26

Slide 28 EPI 1/2 FOV Phase Artifact Slide 29 Phase Contrast Venogram Slide 30 Gradient Instability Artifact 27

Slide 31 Data Acquisition Errors Data corrupted during digitization before FFT (eg electrical transient) Bad memory locations Pattern depends on what data points are modified Slide 32 Data Acquisition Errors Modifying a few points during digitization of a single echo. Slide 33 Signal Corrupted During Scan Similar to a data acquisition error, the signal may be modified by discrete events during any part of the pulse sequence. Their effect upon the image depends upon when and how many events occur. Not at isocenter ν ν 28

Slide 34 Signal Corrupted During Scan Noise appears to be phase incoherent at specific frequencies on upper right. ν Slide 35 Signal Corrupted During Scan Axial ν Sag High frequency or high contrast residual image of leg Two dimensional pattern ν Slide 36 Artifact Summary Artifact vary with imaging conditions Pulse sequence details, coil, patient RF Artifacts External noise phase incoherent, single pixel Other artifacts not along orthogonal directions Phase modulation - ghosts in PE direction Gradient Artifacts Instabilities - Ghosts in PE direction Distortion - direction dependent Severity, appearance direction dependent 29

Slide 37 Acknowledgments Carl Gregory Biomedical Magnetic Resonance Laboratory University of Illinois, Urbana, Illinois http://bmrl.med.uiuc.edu:8080/~cgregory/notebook/artif acts.html 30